ACCURACY OF OSCILLOMETRIC (PETMAP™) AND DOPPLER METHODS TO INDIRECT MEASUREMENT OF BLOOD PRESSURE IN LAMBS

Authors

  • Carla Maria Vela Ulian Universidade Estadual Paulista (UNESP), Botucatu, São Paulo, Brasil
  • Angie Paola Lagos Carvajal Universidade Estadual Paulista (UNESP), Botucatu, São Paulo, Brasil
  • Diana Rocio Becera Velasquez Universidade Estadual Paulista (UNESP), Botucatu, São Paulo, Brasil
  • Francisco José Teixeira Neto Universidade Estadual Paulista (UNESP), Botucatu, São Paulo, Brasil
  • Maria Lúcia Gomes Lourenço Universidade Estadual Paulista (UNESP), Botucatu, São Paulo, Brasil
  • Simone Biagio Chiacchio Universidade Estadual Paulista (UNESP), Botucatu, São Paulo, Brasil

DOI:

https://doi.org/10.1590/1089-6891v17i437301

Keywords:

Animal Sanity

Abstract

Neonatal physiology has peculiarities inherent to the age group. The objective of this study was to monitor the systemic arterial pressure in lambs during neonatal period. We used 20 Ile de France lambs, from birth and at 7, 14, 21, 28 and 35 days of life. The following parameters were analyzed: heart rate (HR), systolic (SBP), diastolic (DBP), and average blood pressure (ABP) by oscillometric method petMAP™, and SBP with Doppler. Invasive pressure validated indirect methods with the average 101.52 ± 12.04 mmHg. The averages with petMAP™ were as follows: HR (156.38 ± 37.46 bpm); DBP (63.80 ± 11.14 mmHg); ABP (81.58 ± 11.83 mmHg); SBP (112.48 ± 15.68 mmHg) and SBP by Dobler (90.27 ± 12.11 mmHg). There was significant difference in HR and blood pressure among the moments. Indirect methods differ between each other in 12.30 mmHg (overestimation   of 11%). Comparing with the invasive methods, both overestimated values of 4% and 16% in   PAS, respectively, for Doppler and petMAP™. The results showed that the Doppler method has established a good relationship with the invasive one being useful for gauging the SBP. The oscillometric method requires larger studies to be used in small ruminants.

Keywords: Doppler; lamb; neonatal period; oscillometric; systemic blood pressure.

Downloads

Download data is not yet available.

Author Biography

Carla Maria Vela Ulian, Universidade Estadual Paulista (UNESP), Botucatu, São Paulo, Brasil

 

 

Published

2016-10-28

How to Cite

ULIAN, Carla Maria Vela; CARVAJAL, Angie Paola Lagos; VELASQUEZ, Diana Rocio Becera; NETO, Francisco José Teixeira; LOURENÇO, Maria Lúcia Gomes; CHIACCHIO, Simone Biagio. ACCURACY OF OSCILLOMETRIC (PETMAP™) AND DOPPLER METHODS TO INDIRECT MEASUREMENT OF BLOOD PRESSURE IN LAMBS. Brazilian Animal Science/ Ciência Animal Brasileira, Goiânia, v. 17, n. 4, p. 593–600, 2016. DOI: 10.1590/1089-6891v17i437301. Disponível em: https://revistas.ufg.br/vet/article/view/37301. Acesso em: 6 dec. 2025.